THPOST029 Upgrade of the Slow Extraction System of the Heidelberg Ion-Beam Therapy Centre’s Synchrotron 2509
 
  • E. Feldmeier, R. Cee, E.C. Cortés García, M. Galonska, Th. Haberer, M. Hun, A. Peters, S. Scheloske, C. Schömers
    HIT, Heidelberg, Germany
 
  The Heidelberg Ion-Beam Therapy Centre HIT consists of a linear accelerator and a synchrotron to provide carbon ions, helium ions and protons for the clinical use as well as oxygen ions for experiments. The RF-KO slow extraction method is used to extract the particles from the synchrotron. To improve the spill quality of the extracted beam a new RF-signal was investigated which increases the R-value from 92.5% to 97,5%. The signal is a multiband RF signal broadened with a random BPSK at 3 frequency bands.  
DOI • reference for this paper ※ https://doi.org/10.18429/JACoW-IPAC2022-THPOST029  
About • Received ※ 07 June 2022 — Revised ※ 13 June 2022 — Accepted ※ 16 June 2022 — Issue date ※ 24 June 2022