JACoW logo

Journals of Accelerator Conferences Website (JACoW)

JACoW is a publisher in Geneva, Switzerland that publishes the proceedings of accelerator conferences held around the world by an international collaboration of editors.


RIS citation export for THPOMS012: Explorative Studies of an Innovative Superconducting Gantry

TY  - CONF
AU  - Pullia, M.G.
AU  - Benedetto, E.
AU  - Dassa, L.
AU  - De Matteis, E.
AU  - Donetti, M.
AU  - Felcini, E.
AU  - Frisella, G.
AU  - Karppinen, M.
AU  - Kurfürst, C.
AU  - Mariotto, S.
AU  - Mereghetti, A.
AU  - Mirandola, A.
AU  - Pella, A.
AU  - Perini, D.
AU  - Piacentini, L.
AU  - Pivi, M.T.F.
AU  - Prioli, M.
AU  - Ratkus, A.
AU  - Rossi, L.
AU  - Sabbatini, L.
AU  - Savazzi, S.
AU  - Stock, M.
AU  - Tommasini, D.
AU  - Torims, T.
AU  - Uberti, S.
AU  - Vannozzi, A.
AU  - Vilcans, J.
AU  - Vretenar, M.
ED  - Zimmermann, Frank
ED  - Tanaka, Hitoshi
ED  - Sudmuang, Porntip
ED  - Klysubun, Prapong
ED  - Sunwong, Prapaiwan
ED  - Chanwattana, Thakonwat
ED  - Petit-Jean-Genaz, Christine
ED  - Schaa, Volker R.W.
TI  - Explorative Studies of an Innovative Superconducting Gantry
J2  - Proc. of IPAC2022, Bangkok, Thailand, 12-17 June 2022
CY  - Bangkok, Thailand
T2  - International Particle Accelerator Conference
T3  - 13
LA  - english
AB  - The Heavy Ion Therapy Research Integration plus (HITRIplus) is a European project that aims to integrate and propel research and technologies related to cancer treatment with heavy ions beams. Among the ambitious goals of the project, a specific work package includes the design of a gantry for carbon ions, based on superconducting magnets. The first milestone to achieve is the choice of the fundamental gantry parameters, namely the beam optics layout, the superconducting magnet technology, and the main user requirements. Starting from a reference 3T design, the collaboration widely explored dozens of possible gantry configurations at 4T, aiming to find the best compromise in terms of footprint, capital cost, and required R&D. We present here a summary of these configurations, underlying the initial correlation between the beam optics, the mechanics, and the main superconducting dipoles design: the bending field (up to 4 T), combined function features (integrated quadrupole), magnet aperture (up to 90 mm), and angular length (30°-45°). The resulting main parameters are then listed, compared, and used to drive the choice of the best gantry layout to be developed in HITRIplus.
PB  - JACoW Publishing
CP  - Geneva, Switzerland
SP  - 2966
EP  - 2969
KW  - dipole
KW  - optics
KW  - superconducting-magnet
KW  - quadrupole
KW  - hadrontherapy
DA  - 2022/07
PY  - 2022
SN  - 2673-5490
SN  - 978-3-95450-227-1
DO  - doi:10.18429/JACoW-IPAC2022-THPOMS012
UR  - https://jacow.org/ipac2022/papers/thpoms012.pdf
ER  -